Feasibility study of a ToF PET prototype with MRPCs

Tamborini, Aurora (2011) Feasibility study of a ToF PET prototype with MRPCs. Il nuovo cimento C, 34 (5). pp. 185-192. ISSN 1826-9885

[img]
Preview
Text
ncc9854.pdf - Published Version

Download (626kB) | Preview
Official URL: https://www.sif.it/riviste/sif/ncc/econtents/2011/...

Abstract

Multigap Resistive Plate Counters (MRPCs) were used to investigate their time resolution to 511 keV gamma rays in view of possible biomedical applications such as Time-of-Flight Positron Emission Tomography (ToF-PET). Using a 22Na source the Time-of-Flight information of annihilation gamma rays has been measured to calculate the MRPC time resolution to 511 keV photons pairs and resulted to be about 145 ps.

Item Type: Article
Uncontrolled Keywords: Computed tomography ; Positron emission tomography (PET)
Subjects: 500 Scienze naturali e Matematica > 530 Fisica
Depositing User: Marina Spanti
Date Deposited: 10 Apr 2020 13:47
Last Modified: 10 Apr 2020 13:47
URI: http://eprints.bice.rm.cnr.it/id/eprint/17480

Actions (login required)

View Item View Item