Tamborini, Aurora (2011) Feasibility study of a ToF PET prototype with MRPCs. Il nuovo cimento C, 34 (5). pp. 185-192. ISSN 1826-9885
|
Text
ncc9854.pdf - Published Version Download (626kB) | Preview |
Official URL: https://www.sif.it/riviste/sif/ncc/econtents/2011/...
Abstract
Multigap Resistive Plate Counters (MRPCs) were used to investigate their time resolution to 511 keV gamma rays in view of possible biomedical applications such as Time-of-Flight Positron Emission Tomography (ToF-PET). Using a 22Na source the Time-of-Flight information of annihilation gamma rays has been measured to calculate the MRPC time resolution to 511 keV photons pairs and resulted to be about 145 ps.