Self-Expandable Transcatheter Aortic Valve Frame Infolding: An Increasingly Recognized Complication

Singh, Gurpreet and Le, Vien and Wiechmann, Robert J and Schreiter, Steven W (2020) Self-Expandable Transcatheter Aortic Valve Frame Infolding: An Increasingly Recognized Complication. European Journal of Case Reports in Internal Medicine, 7 (12). pp. 1-3. ISSN ISSN: 2284-2594

[img]
Preview
Text
2100-Article Text-19049-1-10-20201210.pdf - Published Version

Download (843kB) | Preview
Official URL: https://www.ejcrim.com/index.php/EJCRIM/issue/view...

Abstract

Transcatheter aortic valve replacement (TAVR) with either a balloon-expandable or a self-expandable transcatheter heart valve (THV) is an approved therapy for patients with symptomatic severe aortic stenosis and high or intermediate surgical risk. Here we present a case of severe valve frame infolding of a CoreValve Evolut PRO® self-expandable THV (Medtronic Inc.), which was restored to optimal geometry with balloon post-dilation.

Item Type: Article
Uncontrolled Keywords: Transcatheter aortic valve replacement, cardiac CT scan
Subjects: 600 Tecnologia - Scienze applicate
600 Tecnologia - Scienze applicate > 610 Medicina e salute (Classificare qui la tecnologia dei servizi medici)
600 Tecnologia - Scienze applicate > 610 Medicina e salute (Classificare qui la tecnologia dei servizi medici) > 616 Malattie (classificare qui la Clinica medica, la medicina basata sull'evidenza, la Medicina interna, la Medicina sperimentale)
Depositing User: Marina Spanti
Date Deposited: 20 Dec 2021 14:22
Last Modified: 20 Dec 2021 14:22
URI: http://eprints.bice.rm.cnr.it/id/eprint/21084