La Manna, A. and Abbene, L. and Fauci, F. and Gerardi, G. and Raso, G. and Del Sordo, S. and Barone Tonghi, L. and Stumbo, S. (2006) Portable CdTe detection system for mammographic X-ray spectroscopy. Il nuovo cimento C, 29 (3). pp. 361-367. ISSN 1826-9885
|
Text
ncc9179.pdf - Published Version Download (339kB) | Preview |
Abstract
This paper describes a portable apparatus to be utilized in mammographic X-ray spectroscopy under clinical conditions. The system, based on a CdTe solid-state detector, is able to directly measure mammographic X-ray tube spectra. Good system response to monoenergetic photons was measured using X-ray and γ-ray calibration sources (109Cd and 241Am). The measured molybdenum X-ray spectra, in agreement with simulated spectra, show the good spectral capability of the system also at high photon fluence rates, as typical of clinical mammography. Low tailing, no secondary X-ray escape and low pile-up distortions in the measured spectra indicate that this portable system is suitable for mammographic X-ray spectroscopy.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | X- and gamma-ray spectroscopy ; Solid state detectors ; Quality assurance equipment |
Subjects: | 500 Scienze naturali e Matematica > 530 Fisica |