Portable CdTe detection system for mammographic X-ray spectroscopy

La Manna, A. and Abbene, L. and Fauci, F. and Gerardi, G. and Raso, G. and Del Sordo, S. and Barone Tonghi, L. and Stumbo, S. (2006) Portable CdTe detection system for mammographic X-ray spectroscopy. Il nuovo cimento C, 29 (3). pp. 361-367. ISSN 1826-9885

[img]
Preview
Text
ncc9179.pdf - Published Version

Download (339kB) | Preview
Official URL: https://www.sif.it/riviste/sif/ncc/econtents/2006/...

Abstract

This paper describes a portable apparatus to be utilized in mammographic X-ray spectroscopy under clinical conditions. The system, based on a CdTe solid-state detector, is able to directly measure mammographic X-ray tube spectra. Good system response to monoenergetic photons was measured using X-ray and γ-ray calibration sources (109Cd and 241Am). The measured molybdenum X-ray spectra, in agreement with simulated spectra, show the good spectral capability of the system also at high photon fluence rates, as typical of clinical mammography. Low tailing, no secondary X-ray escape and low pile-up distortions in the measured spectra indicate that this portable system is suitable for mammographic X-ray spectroscopy.

Item Type: Article
Uncontrolled Keywords: X- and gamma-ray spectroscopy ; Solid state detectors ; Quality assurance equipment