Characterization of PVA-GTA Fricke gels dosimeters using MRI and optical techniques in X-rays external radiation therapy

Gallo, S. and Arosio, P. and Locarno, S. and Orsini, F. and Brambilla, M. G. (2021) Characterization of PVA-GTA Fricke gels dosimeters using MRI and optical techniques in X-rays external radiation therapy. Il nuovo cimento C, 44 (4-5). pp. 1-4. ISSN 1826-9885

[img] Text
ncc12290.pdf - Published Version

Download (260kB)
Official URL: https://www.sif.it/riviste/sif/ncc/econtents/2021/...

Abstract

In this work, recent results about the dependence of the response of poly(vinyl-alcohol)-glutaraldehyde Fricke gel dosimeters on the irradiation and holding temperatures are reported. The investigations were carried out by two complementary techniques commonly used in gel dosimetry, namely spectrophotometry and MRI. No significant dependence of the dosimeters sensitivity on the irradiation temperature in the range 20 ◦C–35 ◦C was observed. On the contrary, the holding temperature effects resulted to be not negligible. This work, based on literature results, highlights the limits and the capability of these dosimeters in the 3D dose mapping for clinical practice applications