First characterization of the detector system for the XAFS beam-line of the synchrotron light source SESAME

Cirrincione, D. (2019) First characterization of the detector system for the XAFS beam-line of the synchrotron light source SESAME. Il nuovo cimento C, 42 (5). pp. 1-8. ISSN 1826-9885

[img]
Preview
Text
ncc11938.pdf - Published Version

Download (538kB) | Preview
Official URL: https://www.sif.it/riviste/sif/ncc/econtents/2019/...

Abstract

A Fluorescence Detector System for X-ray Absorption Fine Structure (XAFS), composed of 8 monolithic arrays of Silicon Drift Detectors (SDDs), each having 8 cells of 9 mm2 area, is being realized within the INFN ReDSoX Collaboration. It will be used for X-ray absorption spectroscopy at the Jordanian synchrotron light source SESAME, which provides a photon beam with an energy range between 3 and 30 keV. Detailed characterization tests at Elettra Sincrotrone Trieste have demonstrated an energy resolution at the Mn 5.9 keV Ka line, for the monolithic array, below 170 eV FWHM at room temperature.

Item Type: Article
Subjects: 500 Scienze naturali e Matematica > 530 Fisica
Depositing User: Marina Spanti
Date Deposited: 29 Jan 2021 14:53
Last Modified: 29 Jan 2021 14:53
URI: http://eprints.bice.rm.cnr.it/id/eprint/20678

Actions (login required)

View Item View Item